Back to page

ICS Schneider Messtechnik
ICS Schneider Messtechnik – Kontakt

Dose rate meter GRAETZ GammaTwin

Dose rate meter GRAETZ GammaTwin
  • Small, handy, calibratable dose rate meter for measuring gamma and X-ray radiation for the measurands Ḣ*(10) ambient dose equivalent rate and ambient dose equivalent H*(10) (to record the personal dose, a meter with the measurand Hp(10) must be used).
  • Detector: Energy-compensated Geiger-Müller counter tube
  • Large, digital measured value display on LC display (backlit)
  • Optional simultaneous or separate dose and dose rate display, additional analog display of dose rate as logarithmic bar display
  • 4 factory preset warning thresholds for dose and dose rate; customer-specific modification on request.
  • Switchable acoustic single pulse detection
  • Automatic continuous storage of the dose in a non flüchtiger memory.
  • Storage of the dose and the set parameters even when the battery is changed
  • Constant self-monitoring
  • Designed for rough use, aluminum housing, protection class IP54 (splash-proof)
  • Menu-driven operator guidance
  • Durable nylon belt pouch is included in delivery

Optional:
Test device PV-GTw and test source PS 9 for calibration validity extension to 6 years

Datasheet

Compact electronic dose rate meter with a “twin display” that shows dose rate and dose simultaneously. The dose rate meter GRAETZ GammaTwin is PTB type-tested and therefore calibratable.

 

  • Large backlit LC display with a so-called “twin display”: Simultaneous display of dose rate and dose
  • Acoustic warning signal when warning thresholds are exceeded
  • Intuitive operation
  • Compact and robust design, dust and splash water protection (protection class IP54)

 

Technical data

Display range Dose 0 nSv ≤ H*(10) ≤ 1.0 Sv
Dose rate 0 nSv/h ≤ Ḣ*(10) ≤ 70 mSv/h
Measuring range Dose 0.5 µSv ≤ H*(10) ≤ 1.0 Sv
Dose rate 0.5 µSv/h ≤ Ḣ*(10) ≤ 70 mSv/h
Dimension approx. 28 × 67 × 105 mm
Weight (incl. batteries) approx. 200 g