Portable contamination monitor GRAETZ CoMo-170

- High-performance, mobile contamination monitor with thin-layer plastic scintillation detector (complete elimination of gas-filled or gas-purged detectors).
- A detector can be used to make α-, β- and γ-sensitive measurements, and the measurement system automatically detects whether alpha radiation is present
- Possibility of simultaneous or selective measurement of α- and β-/γ-contaminations
- User-friendly user interface (operation via 5 function keys)
- Settings and measured value parameters protected by password
- Integrated measurement data storage, optional software for reading and processing the measurement data
- Connection option for various external probes, e.g. for dose rate measurement
- Automatic detection of the probes
- Optional stationary use in wall station with power supply
- In case of mechanical destruction of the detector foil, the repair can be carried out inexpensively by the user himself (no costly detector repair in the manufacturer’s factory necessary as with the xenon detector)
- Avoidance of high operating costs for gas flushing
Special versions:
- CoMo-170 ZS and F: adapted to the tasks of the civil defense and the fire department
- CoMo-170 DL: with additional GM counter tube integrated into the front face for measuring the dose rate
- CoMo-170/-300 G: for pure gamma measurements
Datasheet |
The GRAETZ CoMo-170 is a powerful, mobile and battery or accumulator operated contamination monitor for highly sensitive nuclide-related measurement of surfaces for α-, β- and γ-contamination when handling ionizing radiation. The one-hand measuring system has a detector area of 170 cm² and automatically detects whether α-radiation is present. The measured value display is optionally pulses per second (Ip/s) or nuclide-related in Becquerel (Bq) and Bq/cm².
- Easy to use one-hand measuring system
- Fast inspection of large surfaces due to large detector area
- High-sensitivity nuclide-related measurement of surfaces for α-, β- and γ-contamination when handling ionizing radiation
- Dust and splash water protection (protection class IP54)
Technical Data
| Detector type | Thin Film Plastic Scintillation Detector with ZnS Coating | |
| Detector size | 170 cm² | |
| Display ranges | α-channel | up to 5,000 Ip/s |
| β-γ-channel | up to 50,000 Ip/s | |
| Measuring ranges | α-channel | up to 2,500 Ip/s |
| β-γ-channel | up to 20,000 Ip/s | |
| Zero effect | α-channel | approx. 0.1 Ip/s |
| β-γ-channel |
approx. 15 – 20 Ip/s | |
| Dimensions | approx. 280 × 125 × 135 mm | |
| Weight (incl. batteries) |
approx. 800 g | |
| Protection class | IP54 (Protected against splashing water) |
For detailed technical data, please refer to our product data sheet for this article.
Efficiencies for different radionuclides
(Mittelwerte aus Messungen mit 100 cm2 Präparaten) Die nuklidbezogenen Wirkungsgrade des Szintillationsdetektors sind vergleichbar den Wirkungsgraden herkömmlicher Zählrohre.
| C-14 | approx. 14 % | Tc-99m | approx. 3 % |
| F-18 | approx. 18 % | In-111 | approx. 8 % |
| P-32 | approx. 25 % | I-123 | approx. 7 % |
| S-35 | approx. 5 % | I-125 | approx. 12 % |
| Cl-36 | approx. 42 % | I-131 | approx. 21 % |
| K-40 | approx. 30 % | Cs-137 | approx. 35 % |
| Co-57 | approx. 7 % | Au-198 | approx. 23 % |
| Co-60 | approx. 27 % | Tl-204 | approx. 43 % |
| Sr-89 | approx. 27 % | Am-241 α | approx. 22 % |
| Sr-90 / Y-90 (related to Sr-90) | approx. 42 % | P-238 α | approx. 12 % |
| U-238 α | approx. 26 % |
Embodiments
CoMo-170 (version industry)
Item no. 5641 0417-0
CoMo-170 (Medical version)
Art.no.5641 0700-0
CoMo-170 D – with additional GM counter tube for dose rate measurement
Item no. 5641 0696-0
CoMo-170 ZS – Special version for civil defense
Art.no. 5641 0627-0
For CoMo-170 ZS
CoMo-170 ZS – special version for civil defense with accessories
Item no. 5641 0628-0
About the CoMo-170 ZS
CoMo-300 – version with larger detector area of 300 cm².
Item no. 5641 0492-0
CoMo-170 F – special version for the fire brigade*.
Item no. 5641 0487-0
*Different technical data for CoMo-170 F:
Measured value display Ips (not nuclide-related in Bq or Bq/cmÇ)
Software: fire department specific adapted
All functions protected by code word
Additional warning threshold at 3-fold zero effect
No external detectors
No charging function
Accessories