An electronics test bench supplies a device under test located several metres away with 24 V DC. The laboratory power supply is set to exactly 24.0 V. While the device under test receives the expected voltage under no-load conditions, the voltage measured at its input terminals drops significantly as the load current increases. At 12 A, for example, only approximately 22.6 V is available. The power supply continues to display the set voltage and appears to be operating correctly.
The cause often lies not in the voltage regulation of the power supply but in the electrical resistance of the supply cables and connection contacts. Every conductor has a finite resistance. When current flows through this resistance, a voltage drop occurs. This reduces the voltage actually reaching the remote load.
In high-quality laboratory power supplies, the Remote Sense function can largely compensate for this problem within the specified limits. Instead of measuring voltage exclusively at the output terminals of the power supply, two additional sense leads measure the voltage directly at the device under test. The control circuit can then increase the output voltage so that the desired voltage is once again available at the remote load.
Remote Sense is also referred to as remote voltage sensing or external voltage sensing. The function uses a four-wire connection with separate current-carrying supply cables and high-impedance voltage sensing leads. It is essential that the laboratory power supply is specifically designed for this method.
However, compensation has its limits. The power supply requires sufficient voltage and power reserves. Incorrectly connected, reversed or interrupted sense leads can also cause hazardous operating conditions. Particularly with sensitive electronics, it must be considered that incorrect voltage feedback may lead to an undesirable increase in output voltage.
This technical article explains how Remote Sense works and demonstrates how to correctly calculate cable resistance, voltage drops and compensation limits. A practical calculation example compares the load voltage with and without remote sensing. The article also covers the correct location of sense connections, EMC-compliant cable routing, control loop stability, CV/CC operation, overvoltage protection, typical faults and the selection of suitable laboratory power supplies.
Table of Contents
- Define the Power Supply Task and Required Load Voltage
- Understand Why Long Supply Cables Cause Voltage Drops
- Distinguish Remote Sense Regulation from Local Voltage Regulation
- Understand Four-Wire Technology with Force and Sense Connections
- Connect the Sense Leads at the Correct Point on the Load
- Correctly Calculate Cable Resistance, Conductor Cross-Section and Temperature Effects
- Establish the Voltage Budget Between the Power Supply and the Load
- Calculation Example: 24 V at 12 A Over a Five-Metre Cable
- Consider the Maximum Remote Sense Compensation of the Power Supply
- Evaluate Output Voltage, Current Limit and Power Loss Together
- Distinguish Remote Sense Operation in Constant Voltage and Constant Current Modes
- Avoid Incorrect Sense Polarity and Electrical Connection Errors
- Correctly Switch Between Local Sense and Remote Sense
- Route Sense Leads for EMC Compatibility and Avoid Interference
- Safely Evaluate Open Sense Leads and Overvoltage Risks
- Consider Load Steps, Inrush Current and Dynamic Regulation
- Correctly Supply Multiple Loads with Different Voltage Drops
- Check Oscillations, Cable Capacitance and Control Loop Stability
- Verify the Actual Compensation Performance Through Measurement
- Perform a Systematic Testing and Commissioning Procedure
- Diagnose Typical Remote Sense Faults
- Suitable Laboratory Power Supplies and Selection Criteria at ICS Schneider
- Conclusion: Regulate Voltage at the Load While Observing Electrical Limits
- Frequently Asked Questions About Remote Sense on Laboratory Power Supplies
1. Define the Power Supply Task and Required Load Voltage
A laboratory power supply is intended to provide a connected load with a defined electrical voltage and sufficiently high current.
Typical applications include electronics development, industrial test benches, commissioning control units, sensor testing and supplying electrical assemblies.
In many applications, the load is located immediately next to the power supply. The connecting cables are short, and the resulting voltage drops are often small.
The situation is different when the device under test is several metres away or requires high currents.
Even a small electrical resistance can then cause a significant voltage drop.
The first step in designing the power supply is therefore to determine which voltage is actually required at the device under test.
The relevant voltage is not necessarily the voltage present at the output terminals of the power supply.
Particularly with sensitive electronic assemblies, an insufficient supply voltage can cause malfunctions.
Examples include unstable processor supplies, incorrect switching states, undervoltage warnings or unexpected resets of a control unit.
Motors and other dynamic loads may also be sensitive to voltage fluctuations.
Suitable power supply selection therefore requires information about nominal voltage, permissible voltage range, continuous current, peak current, load dynamics and existing cable length.
It must also be determined whether the power supply should regulate the voltage directly at its own output terminals or at the remote load.
Remote Sense is particularly useful when the actual voltage at the load must remain constant within tight limits despite varying load currents.
2. Understand Why Long Supply Cables Cause Voltage Drops
Every electrical conductor has resistance.
This depends, among other factors, on the conductor material, cross-sectional area, cable length and temperature.
When electrical current flows through the conductor, a voltage drop occurs.
According to Ohm’s law:
ΔU = I · R
Here, I denotes current and R represents electrical resistance.
For a DC power supply, the entire current-carrying path must be considered.
This means that both the positive and negative supply conductors contribute to the total voltage drop.
If a load is located five metres from the power supply, the total conductor length in a simple two-wire connection is ten metres.
Additional contact resistances may also occur.
These include connectors, terminals, fuses, switches and relay contacts.
Such resistances can have a significant effect, particularly at higher currents.
An additional resistance of 20 mΩ, for example, causes a voltage drop of 0.2 V at 10 A.
At 20 A, this voltage drop increases to 0.4 V.
The voltage at the load can be calculated approximately as:
Uload = Upower supply − ΔUtotal
It is particularly important to understand that a regulated power supply operating with local voltage sensing does not automatically compensate for the voltage drop in the external cables.
The instrument may maintain exactly the set voltage at its own output terminals while the voltage at the load is significantly lower.
3. Distinguish Remote Sense Regulation from Local Voltage Regulation
A regulated laboratory power supply requires feedback about the actual output voltage.
In normal local regulation mode, the voltage is measured inside the power supply or directly at its output terminals.
The controller compares this measured voltage with the setpoint.
If a deviation occurs, the instrument adjusts its output voltage accordingly.
This method is called Local Sense.
Losses occurring in longer external supply cables are not fully detected by this arrangement.
With Remote Sense, the voltage feedback point is moved to the remote load.
Two additional leads measure the voltage directly at the intended load.
The internal sense inputs normally have high electrical input impedance.
Consequently, only a very small current flows through the sense leads.
The voltage drop along these leads is therefore correspondingly small under normal operating conditions.
The power supply thus receives feedback about the actual supply voltage at the device under test.
If the load voltage is too low, the power supply can increase its output voltage.
This adjustment takes place automatically and continuously during normal Remote Sense operation.
For example, a power supply with a setpoint of 24.0 V may increase its output voltage to 24.8 V if the total voltage drop across the supply cables is 0.8 V.
Approximately 24.0 V is then available again at the load.
The cable itself does not become resistance-free.
The voltage drop and associated power dissipation remain physically present.
Remote Sense compensates for voltage loss by increasing the power supply’s output voltage accordingly.
4. Understand Four-Wire Technology with Force and Sense Connections
A Remote Sense connection normally uses four electrical conductors.
Two conductors carry the actual load current.
The other two are used exclusively for voltage feedback.
The current-carrying connections are frequently labelled Output + and Output − or Force + and Force −.
The additional voltage sensing connections may be labelled Sense + and Sense −.
Designations can vary depending on the manufacturer.
However, the basic assignment of functions remains the same.
| Connection | Function | Current Loading | Important Requirement |
|---|---|---|---|
| Output + / Force + | Positive power supply connection to the load | Carries load current | Sufficient conductor cross-section and suitable current-carrying capacity |
| Output − / Force − | Negative return conductor of the power supply | Carries load current | Sufficient conductor cross-section and reliable return path |
| Sense + | Measures the positive voltage potential at the load | Normally only a very small input current | Correct polarity and reliable contact |
| Sense − | Measures the negative voltage potential at the load | Normally only a very small input current | Correct polarity and reliable contact |
The actual power supply continues to operate through the two Force conductors.
The Sense leads must not be used as an additional power current path.
It is particularly important not to confuse the voltage sensing lead with the normal current return conductor.
In four-wire technology, power transmission and voltage measurement are functionally separated.
The method resembles Kelvin measurement technology, in which current and voltage connections are separated to account for lead resistance during precision measurements.
However, unlike a pure four-wire resistance measurement, the laboratory power supply actively uses the detected voltage to regulate its output.
This requires the power supply to have suitable Sense inputs and a regulation circuit specifically designed for this function.
5. Connect the Sense Leads at the Correct Point on the Load
The position of the voltage sensing point determines which cable losses are actually compensated.
Ideally, the Sense leads are connected directly to the designated supply terminals of the device under test.
The power supply can then measure the voltage actually present at the load.
If Sense + and Sense − are instead connected at the output terminals of the power supply, the downstream cable section is not compensated.
The instrument then effectively operates with local voltage feedback.
A connection at an intermediate terminal may also be insufficient.
If additional cables, connectors or switching devices are located between this terminal and the actual load, their voltage drops are not fully detected.
Voltage feedback must therefore be taken from the point where the voltage is intended to remain constant.
On an electronics test bench, this is frequently the supply input of the device under test.
In larger installations, a specifically defined distribution point may also serve as the regulation reference.
The appropriate position depends on the electrical installation structure.
Reliable contact points are particularly important.
Loose Sense connections can alter or interrupt voltage feedback.
Mechanical tensile loads on Sense leads should also be avoided.
Correct placement of the voltage sensing connection is therefore not only a matter of measurement accuracy but also of operational safety.
6. Correctly Calculate Cable Resistance, Conductor Cross-Section and Temperature Effects
The electrical resistance of a conductor depends on its material, length and cross-sectional area.
For a simplified calculation:
R = ρ · l / A
Where:
R is the electrical conductor resistance, ρ is the electrical resistivity of the material, l is the conductor length and A is the conductor cross-sectional area.
For copper, a resistivity of approximately 0.0178 Ω·mm²/m can be used for an approximate calculation at 20 °C.
For a two-wire DC power supply, the complete outgoing and return conductor path must be considered.
If both conductors have the same length and cross-sectional area, the following approximation applies:
Rtotal = ρ · (2 · l) / A
Here, l represents the one-way distance between the power supply and the load.
Initially, the formula considers only the copper conductors.
Additional contact, fuse and connection resistances must be included where necessary.
Temperature is also relevant.
The electrical resistance of copper increases as temperature rises.
If a conductor heats up due to load current, its voltage drop may also increase.
Remote Sense can compensate for such changes within its permissible regulation limits.
However, the resulting heat generation is not eliminated.
Remote Sense must therefore not be considered a substitute for adequate cable sizing.
Conductor cross-section, current-carrying capacity, installation method, ambient temperature and electrical protection measures must still be suitable for the application.
7. Establish the Voltage Budget Between the Power Supply and the Load
For reliable Remote Sense design, it is necessary to determine how much additional output voltage is required.
For a simple DC supply with resistive cable losses:
Upower supply = Uload + I · Rtotal
Without Remote Sense, the power supply is normally set to a specific output voltage.
The load then receives this output voltage minus the voltage drop along the cable.
With Remote Sense, the controller attempts to maintain the voltage directly at the load at the specified setpoint.
To achieve this, the instrument must provide a correspondingly higher voltage at its output terminals.
For example, at a load current of 10 A and a total cable resistance of 0.08 Ω:
ΔU = 10 A · 0.08 Ω = 0.8 V
To provide the required 24.0 V at the load, the power supply would need to output approximately 24.8 V under these idealised conditions.
The required compensation is 0.8 V.
It is important to note that load current is not always constant.
With an electronic load, current may vary between standby operation, normal operation and brief peak-load conditions.
The voltage drop therefore changes as well.
In constant voltage operation, Remote Sense is intended to regulate these changes within its specified static and dynamic limits.
For a reliable design, both the normal continuous current and relevant peak currents must therefore be considered, together with the maximum available compensation voltage.
8. Calculation Example: 24 V at 12 A Over a Five-Metre Cable
An electronic device under test must be supplied with 24.0 V DC.
The device under test is located five metres from the laboratory power supply.
The positive and negative supply conductors are made of copper, each with a cross-sectional area of 1.5 mm².
A conductor temperature of 20 °C is assumed for the idealised calculation.
The load current can reach 12 A.
For simplicity, an electronic load with a defined current is considered.
The total current path is:
ltotal = 5 m + 5 m = 10 m
Using the electrical resistivity of copper:
Rtotal = 0.0178 Ω·mm²/m · 10 m / 1.5 mm²
Rtotal ≈ 0.1187 Ω
At 12 A, the voltage drop is:
ΔU = 12 A · 0.1187 Ω ≈ 1.424 V
Without Remote Sense, with a regulated power supply output of 24.0 V, only approximately 22.576 V would reach the device under test under these simplified conditions.
Uload = 24.0 V − 1.424 V = 22.576 V
With a suitable Remote Sense function, the power supply can attempt to maintain 24.0 V at the device under test.
To do so, it would need to provide approximately 25.424 V at its output terminals.
Upower supply,remote = 24.0 V + 1.424 V = 25.424 V
The following table shows the behaviour at different load currents.
| Load Current | Cable Voltage Drop | Load Voltage Without Sense | Required Power Supply Output Voltage with Sense | Cable Power Loss |
|---|---|---|---|---|
| 0 A | 0.000 V | 24.000 V | 24.000 V | 0.00 W |
| 3 A | 0.356 V | 23.644 V | 24.356 V | 1.07 W |
| 6 A | 0.712 V | 23.288 V | 24.712 V | 4.27 W |
| 9 A | 1.068 V | 22.932 V | 25.068 V | 9.61 W |
| 12 A | 1.424 V | 22.576 V | 25.424 V | 17.09 W |
The example demonstrates the main benefit of Remote Sense.
As current increases, the voltage drop increases.
Without appropriate compensation, the voltage supplied to the device under test decreases.
With suitable remote sensing, the load voltage can be maintained at an almost constant value within the instrument’s operating limits.
However, the table also reveals another important relationship.
At 12 A, the power dissipated in the cable already reaches approximately 17.1 W.
For resistive cable losses:
Ploss = I² · R
Ploss = (12 A)² · 0.1187 Ω ≈ 17.1 W
This power is converted into heat along the two current-carrying copper conductors.
Remote Sense does not eliminate this heat generation.
Furthermore, the calculation must not be interpreted as general approval to operate a 1.5 mm² conductor at 12 A under arbitrary installation conditions.
The actual current-carrying capacity, installation method, permissible temperature rise, protection and cable requirements must be evaluated separately.
The calculated compensation also assumes that the power supply is actually permitted to provide the required output voltage at 12 A and has a sufficiently large Remote Sense compensation range.
A power supply with a maximum output voltage of 24 V, for example, could not provide the calculated 25.424 V.
Even a power supply with a maximum output voltage of 30 V would only be suitable if its Remote Sense function and permissible power rating explicitly support the required operating condition.
The values are a physical calculation example and do not represent the specifications of any particular power supply offered by ICS.
9. Consider the Maximum Remote Sense Compensation of the Power Supply
Remote Sense does not provide unlimited compensation.
Every suitably equipped power supply has its own design and electrical operating limits.
These include the maximum permissible output voltage, current limit, output power and, where applicable, a separately specified maximum sense compensation voltage.
A power supply may, for example, be set to 24 V but compensate for only a limited additional voltage drop across the output cables.
The size of this range depends on the particular model.
For various DC power supplies, Keysight specifies a limited permissible compensation for voltage drops in individual load leads, for example.
Such specifications are instrument-specific and must not be applied indiscriminately to other power supplies.
For correct design, the desired load voltage is therefore not the only important parameter.
The maximum required output voltage must also be determined:
Urequired,max = Uload,setpoint + ΔUcable,max
Additional voltage drops across connectors or other series components must also be included.
If the required compensation exceeds the permissible range of the power supply, the load voltage may fall below the setpoint despite Remote Sense being enabled.
Depending on the instrument configuration, the power supply may then reach an output voltage, power or other regulation limit.
The electrical limit may even be reached before the calculated maximum output voltage is exhausted.
For example, a separately specified Sense compensation limit may be smaller than the remaining difference to the maximum output voltage.
For demanding applications, the actual specified maximum compensation of the selected instrument must therefore be verified.
10. Evaluate Output Voltage, Current Limit and Power Loss Together
A higher output voltage at the same load current means that the power supply must deliver correspondingly more power.
For DC electrical power:
P = U · I
If the device under test requires 24 V at 12 A, its electrical power consumption is:
Pload = 24 V · 12 A = 288 W
In the calculation example, the power supply must provide approximately 25.424 V at 12 A because of cable losses.
This results in an output power of:
Ppower supply ≈ 25.424 V · 12 A ≈ 305.1 W
The difference of approximately 17.1 W corresponds to the calculated cable power loss.
This additional power requirement can be relevant, particularly with laboratory power supplies that have a limited maximum output power.
An instrument with sufficient nominal voltage and current ratings is not automatically suitable for every combination of these values.
Some power supplies have additional power characteristics that limit their permissible operating range.
The current limit must also be suitable for the load.
If the load demands more current than permitted, the power supply may change from constant voltage to constant current operation.
The desired load voltage may then no longer be maintained.
Reliable selection must therefore consider output voltage, continuous current, possible current peaks, output power and the additional compensation requirement together.
11. Distinguish Remote Sense Operation in Constant Voltage and Constant Current Modes
Regulated laboratory power supplies frequently provide constant voltage and constant current operation.
In constant voltage mode, or CV, the power supply attempts to maintain the configured voltage at a constant value.
Remote Sense allows this voltage regulation loop to be referenced to the remote load.
The power supply then regulates the voltage measured at that point to the specified setpoint.
In constant current mode, or CC, the current is limited or regulated to a defined value instead.
The output voltage is then determined by the electrical load and the permissible operating limits.
In this condition, a Remote Sense function intended for voltage regulation cannot force the load voltage to reach the CV setpoint simultaneously.
The actual handling of Sense inputs in CC mode is instrument-specific.
Some manufacturers explicitly describe Remote Sense as a function for CV operation.
With certain instruments, using the Sense inputs in other operating modes may be subject to restrictions.
The manufacturer’s instructions must therefore be followed.
Diagnosis at high load currents is particularly important.
If the load voltage falls below the setpoint despite Remote Sense being enabled, the power supply may already be operating in constant current mode.
In this case, the Sense wiring is not necessarily incorrect.
The load may be demanding more current than the configured or technically permissible current limit allows.
The indication of the actual operating mode can therefore provide an important diagnostic clue.
The general relationship between CV and CC is explained in greater detail in the ICS technical article Laboratory Power Supply in CV/CC Mode: Understanding Current Limiting and the Transition Between Constant Voltage and Constant Current.
12. Avoid Incorrect Sense Polarity and Electrical Connection Errors
The Sense connections must be connected with the correct polarity.
Sense + belongs to the positive voltage potential of the load.
Sense − belongs to the negative voltage potential of the load.
Reversing these connections can significantly impair the electrical feedback of the power supply.
Depending on the instrument configuration, protective functions may respond or impermissible voltage conditions may occur.
Damage to the Sense inputs is also possible under certain incorrect connection conditions.
It is particularly important that Sense leads are not connected to arbitrary external electrical potentials.
The permissible voltages between Sense inputs, output terminals and, where applicable, protective earth are manufacturer-specific.
Additional restrictions may apply to non-isolated power supplies or supplies with a shared electrical reference.
With multichannel power supplies, the Sense leads must also be assigned to the correct output channel.
Cross-connecting different channels is not automatically permissible.
The actual terminal assignment must always be taken from the operating instructions.
The complete wiring should be checked before activating the output voltage for the first time.
Sensitive devices under test in particular may be endangered by incorrect voltage feedback.
13. Correctly Switch Between Local Sense and Remote Sense
Not every power supply uses the same technical method for selecting between local and remote voltage regulation.
On some instruments, the Sense terminals are connected to the output terminals in the factory configuration.
Voltage regulation then measures the voltage directly at the power supply.
To use Remote Sense on these instruments, the designated local connections must be changed according to the manufacturer’s instructions.
Other power supplies provide a menu setting or automatic detection of an external Sense connection.
In certain configurations, the required feedback can also be switched internally between Local Sense and Remote Sense.
Consequently, one universal connection procedure must not be assumed for all instruments.
Removing existing jumper links without understanding the specific power supply function could be particularly critical.
An incorrect configuration may also result in connected Sense leads while the instrument continues regulating only the voltage at its own output terminals.
The actual operating method must be established from the manufacturer documentation.
During commissioning, switching between regulation modes should take place in the safe condition specified for the instrument.
Particularly with sensitive electronic loads, undesirable voltage transients during switching must be prevented.
Successful activation of Remote Sense can subsequently be verified using suitable voltage measurements directly at the load.
14. Route Sense Leads for EMC Compatibility and Avoid Interference
Sense leads are used for voltage measurement rather than carrying high load currents.
Their electrical input impedance is normally high.
However, this can also make them sensitive to certain forms of electromagnetic interference.
Industrial test benches frequently contain contactors, variable frequency drives, motor cables and other rapidly switching electrical components nearby.
These may induce interference voltages in improperly routed measuring leads.
If the voltage control circuit interprets such interference as actual changes in load voltage, regulation behaviour may be affected.
Cable routing appropriate to the power supply is therefore important.
Depending on the manufacturer and instrument architecture, recommendations may include twisted Sense lead pairs, suitable shielding or a defined physical arrangement relative to the power cables.
The exact recommendations may differ between power supplies.
With some instruments, the Sense leads should be routed together with the associated power cables in a specified manner to avoid undesirable regulation dynamics.
A general rule stating that “Sense leads must always be routed separately from supply cables” would therefore not be technically correct for every power supply.
The actual cable routing must comply with the respective manufacturer instructions.
It is particularly important that Sense leads are not routed immediately next to strong interference sources if this could cause unacceptable interference coupling.
The shielding concept is also instrument-specific.
Unsuitable shield connections may increase interference or create undesirable electrical potential connections.
A complete EMC evaluation of the power supply and measurement arrangement is therefore advisable for demanding test benches.
15. Safely Evaluate Open Sense Leads and Overvoltage Risks
An interrupted Sense lead can be a particularly critical fault condition.
In a genuine Remote Sense system, the voltage sensing connections form part of the feedback path of the voltage regulation loop.
If this feedback path is interrupted, the controller may detect an incorrect voltage value.
In an unsuitable or unprotected system, the power supply might attempt to compensate for a supposedly low load voltage by increasing its output voltage.
This could result in dangerous overvoltage at the load.
Manufacturers of high-quality power supplies therefore sometimes use special protective circuits for interrupted Sense leads.
Depending on the instrument configuration, these may enable a return to local voltage regulation or detect a corresponding fault.
Other instruments have dedicated open-sense detection that switches off the output under certain fault conditions.
The actual behaviour must be evaluated using the manufacturer documentation.
It must not be assumed that every laboratory power supply automatically enters a safe state when a Sense connection is interrupted.
Overvoltage protection is therefore particularly important.
Many suitable power supplies feature adjustable Overvoltage Protection, or OVP.
This can help protect the device under test against certain impermissible output voltages.
The correct setting must be matched to the actual Remote Sense application.
On many power supplies, the OVP threshold is monitored at the output terminals.
Because the output voltage in Remote Sense mode can be higher than the desired load voltage, this normal compensation voltage must be considered when setting the protection threshold.
However, the OVP threshold must not be set arbitrarily high.
The maximum permissible voltage of the device under test must also be considered.
Additional protection directly at the load may be useful, depending on the application, provided that it is compatible with the complete power supply system.
The specific protective functions, Sense lead monitoring and permissible fault responses are therefore important selection criteria.
16. Consider Load Steps, Inrush Current and Dynamic Regulation
An electronic load does not always require a constant current.
For example, the current consumed by a processor assembly can change rapidly between different operating states.
Motors, heaters and power electronics can also produce significant current variations.
When the current changes rapidly, the voltage drop across the supply cable changes as well.
Remote Sense attempts to compensate for this variation by adjusting the output voltage of the power supply.
However, the controller requires a certain response time.
Very rapid load steps may therefore cause short-term voltage deviations at the load.
In addition to resistive properties, the connecting cable also has electrical inductance.
Rapid current changes can therefore produce an additional dynamic voltage component.
For a simple inductive analysis:
UL = L · di/dt
This relationship demonstrates that fast current changes can produce significant transient voltages even when the resistive cable loss is relatively small.
Remote Sense regulation cannot necessarily compensate for these effects instantaneously.
Suitable local decoupling directly at the device under test may therefore be required, particularly with sensitive digital electronics.
The selection of such decoupling capacitors must be compatible with the power supply control loop.
An arbitrarily large capacitance is not automatically advantageous.
The output dynamics and permissible additional load capacitance must also be considered.
For test benches with rapid load changes, the dynamic behaviour of the complete power supply path should therefore be investigated.
17. Correctly Supply Multiple Loads with Different Voltage Drops
Another important question arises when one laboratory power supply is intended to supply several loads simultaneously.
For example, a common DC supply may power various electronic assemblies within a test bench.
The loads may be located at different positions and connected using cables of different lengths.
This can result in different voltage drops.
However, a single Remote Sense control loop can normally only measure the voltage between one defined Sense + and Sense − connection point.
If the voltage sensing connections are located directly at load A, for example, the power supply regulates the voltage at that point.
Load B may still receive a lower voltage because of resistance in its own connecting cables.
Different load currents can cause further variations in voltage drop.
The Sense leads must therefore not be connected arbitrarily to different loads without a suitable overall design.
A properly dimensioned common DC distribution point is often a useful solution.
With a suitable power supply, the voltage can then be regulated at this defined point.
Any remaining voltage drops between the distribution point and individual loads must be considered separately.
If high individual voltage accuracy is required, separate supply channels or suitable local voltage regulators may be necessary.
With multichannel laboratory power supplies, the Sense inputs must be correctly assigned to the respective channels.
A common Sense connection across several independent outputs is not automatically permissible.
The instrument architecture and manufacturer specifications determine which configurations are supported.
18. Check Oscillations, Cable Capacitance and Control Loop Stability
Remote Sense changes the effective feedback path of the voltage regulation loop.
Voltage measurement no longer takes place directly at the power supply output but at the remote load.
The electrical characteristics of the connecting cables therefore become part of the control loop.
Particularly long cables can affect dynamic behaviour because of their inductance, capacitance and resistance.
The load may also contain input capacitors or its own switching regulators.
These can interact with the power supply regulation.
Under unfavourable conditions, the voltage control loop may oscillate or generate undesirable overshoots.
A typical symptom is a load voltage that exhibits repeated small or large fluctuations despite an apparently stable setting.
Such effects can be particularly noticeable during rapid load changes.
Manufacturers of suitable power supplies therefore sometimes specify requirements for Sense lead routing, maximum cable lengths and permissible additional capacitances.
Additional damping measures may be necessary in certain applications.
However, such measures must not be applied arbitrarily.
In particular, an additional capacitor directly at the load may reduce certain rapid voltage variations, but it also affects load current, inrush current and control loop stability.
Reliable design therefore requires the actual power supply, connecting cable and device under test to be considered together.
For demanding applications, a measurement-based investigation using a suitable oscilloscope and correctly designed test setup is advisable.
A conventional digital multimeter may not adequately indicate rapid transient events.
It is particularly important to distinguish between a stable average DC voltage and a genuinely low-noise supply under dynamic operating conditions.
19. Verify the Actual Compensation Performance Through Measurement
Whether Remote Sense is functioning should not be assessed solely by checking that a menu option is enabled.
The decisive factor is the voltage actually present at the device under test.
For suitable verification, the voltage at the power supply output can be compared with the voltage directly at the supply terminals of the device under test.
An independent voltage reference or suitable digital multimeter can be used for this purpose.
Measurements should be performed under defined load conditions.
In local regulation mode, the power supply may provide the set voltage at its output terminals while the device under test receives a lower voltage because of cable losses.
With correctly functioning Remote Sense regulation, the load voltage should be closer to the setpoint within the specified instrument limits.
At the same time, the power supply output voltage increases in accordance with the cable voltage drop.
A comparison at several load currents is particularly informative.
This makes it possible to determine how the voltage drop changes and whether the regulation correctly compensates within its intended operating range.
Voltages should also be examined at the highest intended load currents.
When the compensation limit is reached, the load voltage may again fall below the setpoint.
The actual CV/CC operating condition should also be documented.
With dynamic loads, the voltage waveforms over time may be important as well.
Accurate measurements require reference measuring instruments with sufficiently low measurement uncertainty.
The location of the measurement probes is also relevant.
Measuring voltage at an intermediate terminal is not necessarily equivalent to measuring it directly at the input of the device under test.
20. Perform a Systematic Testing and Commissioning Procedure
Remote Sense should only be used if the designated power supply explicitly supports this function.
The following test sequence describes a general technical approach. Actual wiring and activation must comply with the manufacturer instructions and the safety requirements of the test bench.
- Define the measurement task: Record nominal voltage, permissible voltage range, continuous current, peak loads and voltage stability requirements.
- Identify the power supply: Check whether the laboratory power supply supports genuine Remote Sense functionality.
- Check the Sense specifications: Determine terminal assignment, maximum compensation voltage, permissible cable length and protective functions.
- Record the supply path: Document the distance to the load, conductor cross-section, material and existing connection components.
- Calculate cable resistance: Include positive and negative supply conductors and relevant contact resistances.
- Determine the maximum voltage drop: Use the highest intended load current and, where necessary, additional thermal influences.
- Check the voltage margin: Ensure that the power supply is permitted to provide the required increased output voltage.
- Evaluate power and current limits: Check output current, power limits and the required load reserve.
- Check cable sizing: Evaluate permissible current-carrying capacity, heating, protection and mechanical suitability.
- Define the Sense connection point: Determine the actual point at which voltage is to be regulated.
- Plan cable routing: Consider manufacturer requirements for Sense leads, EMC, shielding and cable routing.
- Check polarity: Verify the correct assignment of Force +, Force −, Sense + and Sense −.
- Define Local/Remote configuration: Consider the operating mode and any switching procedure specified by the manufacturer.
- Design overvoltage protection: Match OVP functionality and any additional protective measures to the actual application.
- Establish the connection under safe conditions: Connect the cables according to the procedure specified for the power supply.
- Check initial operating conditions: Verify output voltage and load voltage under suitable defined conditions.
- Verify Remote Sense: After activation according to the manufacturer’s instructions, check the load voltage and the necessary adjustment of the power supply output voltage.
- Test the load range: Investigate relevant load currents, including the maximum intended operating current.
- Evaluate dynamic behaviour: Where appropriate, investigate load steps, start-up conditions and possible control loop oscillations.
- Document the results: Record the instrument type, configuration, cables, load conditions, voltage readings, protective functions and any deviations identified.
It is particularly important not to perform testing exclusively under no-load conditions.
Without significant load current, there is normally no relevant resistive voltage drop.
The actual compensation function can therefore be evaluated particularly effectively at defined higher currents.
Supplying a sensitive device under test must also involve suitable current limiting and the required overvoltage protection.
Uncontrolled disconnection or reconnection of Sense leads during operation is not a suitable general testing procedure.
Permissible testing of fault conditions is instrument-specific and may only be performed according to an appropriate safe procedure.
21. Diagnose Typical Remote Sense Faults
Remote Sense problems may result from insufficient voltage margin, incorrect connections or unfavourable regulation conditions.
The following overview shows typical observations and possible technical causes.
| Observation | Possible Cause | Suitable Check |
|---|---|---|
| Load voltage continues to decrease as current increases | Remote Sense not enabled, incorrect sensing point or compensation limit reached | Check operating mode, Sense connections and available output voltage |
| Sense leads connected, but no noticeable voltage compensation | Power supply does not support genuine remote sensing or remains in Local Sense mode | Check instrument functions and manufacturer documentation |
| Power supply indicates a higher voltage than the configured load setpoint | Normal compensation for the cable voltage drop | Measure voltage directly at the device under test and compare with the setpoint |
| Output voltage increases unexpectedly and significantly | Interrupted or incorrectly connected Sense lead, faulty feedback | Safely stop the test and have the Sense connections and protective functions checked |
| Power supply reports a Sense fault | Open lead, impermissible Sense voltage or manufacturer-specific diagnostic condition | Evaluate the fault code and permissible connection conditions using the operating instructions |
| Load voltage fluctuates or oscillates | Control loop instability due to cable impedance, unsuitable routing or load capacitance | Investigate the cable arrangement and time-dependent voltage waveform |
| Load voltage falls below the setpoint despite Remote Sense, and CC is indicated | Current limit reached | Check load current, current limit and permissible operating range |
| Remote Sense works at low loads but not at high current | Voltage, current, power or compensation limit reached | Check the complete voltage and power budget at maximum load current |
| Only one of several loads receives the desired voltage | Sense regulation refers to a single sensing point | Evaluate the distribution concept and voltage drops to the other loads |
| Power supply shuts down after overvoltage protection is activated | OVP threshold does not adequately account for the normal increased output voltage | Check the OVP sensing point, compensation requirement and permissible instrument specifications |
| Voltage is correct under steady-state conditions but unstable during load steps | Limited regulation speed, cable inductance or unsuitable decoupling | Investigate transients using suitable time-resolved measuring equipment |
| Required compensation changes after replacing the supply cable | Different cable resistance or changed contact resistance | Recheck cable specifications, connection points and load voltage |
The table describes possible causes and suitable diagnostic approaches.
A definitive fault diagnosis must be performed on the specific power supply and the complete supply circuit.
It is particularly important to distinguish between a normally increased output voltage during Remote Sense operation and a dangerous regulation fault.
The intended load voltage, actual electrical operating conditions and protective functions of the power supply are decisive.
22. Suitable Laboratory Power Supplies and Selection Criteria at ICS Schneider
22.1 Remote Sense as a Mandatory Selection Criterion
For the active remote compensation described in this article, the selected laboratory power supply must explicitly provide a suitable Remote Sense function.
A specification indicating a regulated output voltage or precise CV regulation is not sufficient.
Likewise, a remote control input and a monitor output are not automatically equivalent to Remote Sense.
A remote control input can be used to adjust voltage or current externally, for example.
A monitor output can provide electrical operating values for external indication.
With Remote Sense, however, the voltage actually measured at the remote load must be used within the intended regulation loop to adjust the output voltage automatically.
Suitable selection therefore requires confirmation of genuine Sense connections or the corresponding technical function.
The following instruments are regulated laboratory power supplies listed in the ICS product range. Where Remote Sense is not explicitly specified, this function must not be assumed to be available on the respective instrument.
22.2 2258.1 High-Power DC Voltage Regulator: DC Supply up to 20 A
The 2258.1 High-Power DC Voltage Regulator is a regulated laboratory power supply for industrial testing and power supply applications.
The instrument provides an adjustable output voltage range of 0 … 48 V DC.
The maximum output current is 20 A.
The specified output power is 960 W.
For regulation, the instrument uses a linear regulator with triac pre-regulation.
It provides automatic switching between constant voltage and constant current modes.
Overtemperature protection and short-circuit protection are also specified.
Load regulation in CV mode is specified as 30 mV for a load change from 0 to 100 %.
ICS specifies an RMS ripple voltage of 3 mV in CV mode.
The instrument may be suitable for test benches requiring high DC output power.
With a remote load, the voltage drop across the supply cables must be considered during system design.
The ICS product information states that a remote control input and monitor output are available on request.
However, a genuine Remote Sense function is not explicitly confirmed in the published product information reviewed.
A corresponding configuration or suitability must therefore be clarified separately before planning active remote voltage compensation.
22.3 2257.1 High-Power DC Voltage Regulator: Regulated High-Current Supply up to 40 A
The 2257.1 high-power DC voltage regulator listed by ICS provides an output voltage range of 0 … 30 V DC and a maximum output current of 40 A.
The documented output power is 1,200 W.
The instrument uses linear regulation with triac pre-regulation.
Its functions include constant voltage and constant current regulation, overtemperature protection and short-circuit protection.
The specified CV load regulation is 50 mV.
ICS specifies an RMS ripple voltage of 1 mV in CV mode.
Voltage drops across external connecting cables become particularly relevant at high output currents.
With a total cable resistance of only 20 mΩ, a current of 40 A already produces a voltage drop of 0.8 V.
The corresponding cable power loss is 32 W.
These values demonstrate why conductor cross-section, current-carrying capacity and contact resistance must be carefully evaluated for high-current power supplies.
An optional remote control input and monitor output are also described for this instrument.
However, Remote Sense is not listed as an available function in the ICS product information reviewed.
Active remote compensation must therefore not be assumed without corresponding manufacturer confirmation.
Further information about the product group is available in the High-Power Bench Units category.
22.4 2258.7 High-Power DC Voltage Regulator: 30 V DC at up to 30 A
The 2258.7 is another high-power DC voltage regulator in the ICS product range.
Its documented output specifications are 0 … 30 V DC and 0 … 30 A.
The instrument provides regulated voltage and current output, automatic CV/CC switching and corresponding protective functions.
The manufacturer describes the design as a bench unit with active cooling.
According to the published product information, CV load regulation is 60 mV.
RMS ripple voltage is specified as 2 mV.
The instrument may be suitable for larger electronic devices under test and industrial test setups.
With a physically remote load, possible cable losses and the available output voltage margin must be considered.
An optional remote control input or monitor output is not equivalent to automatic voltage compensation using Remote Sense.
As remote voltage sensing is not confirmed in the published product information reviewed, its availability or an appropriately suitable alternative must be clarified explicitly.
22.5 2229.5 DC Voltage Regulator: Multiple Outputs for Laboratory and Development Applications
The 2229.5 DC Voltage Regulator is a laboratory power supply with multiple output voltages.
It provides two adjustable DC outputs, each covering 0 … 32 V DC at up to 3 A.
An additional fixed-voltage output provides 5 V at 3 A.
Each of the two variable outputs has a power rating of 96 W.
The instrument uses linear voltage and current regulation.
Automatic switching between CV and CC operation is also included in the documented functions.
The published CV load regulation is 40 mV for a load change from 0 to 100 %.
RMS ripple voltage is specified as 2 mV.
The 2229.5 may be suitable for laboratory experiments, electronics development and smaller test setups.
When supplying several loads, their different cable resistances and load currents must be considered.
A Remote Sense function is not explicitly mentioned for this model in the published product information reviewed.
Such functionality must therefore not be inferred solely from the regulated output voltage or multichannel design.
22.6 Selecting a Laboratory Power Supply with Genuine Remote Sense Functionality
If Remote Sense is mandatory for an application, this requirement must be clearly specified during the technical enquiry stage.
In addition to the required load voltage and maximum output current, the additional compensation voltage required and the length of the supply cables are particularly important.
The maximum permissible output power must also be suitable for the application.
Protective functions should be evaluated carefully, particularly with sensitive electronic devices under test.
These include suitable overvoltage protection, handling of interrupted Sense leads and behaviour in the event of a voltage feedback fault.
The permissible conditions for cable length, EMC, dynamic load changes and, where applicable, multichannel operation must also be considered.
Suitable manufacturer documentation confirming the Remote Sense function is therefore more important than general descriptions such as “precision laboratory power supply” or “remotely controllable power supply”.
ICS Schneider can assist in selecting and technically clarifying suitable laboratory power supplies.
An overview of the listed instruments is available in the Laboratory Power Supplies and Power Supplies categories.
23. Conclusion: Regulate Voltage at the Load While Observing Electrical Limits
Remote Sense is an effective function for improving voltage regulation when supplying remote electrical loads.
The method uses two additional voltage sensing leads to measure the voltage actually present at the device under test.
A suitably equipped power supply can then increase its output voltage and compensate for voltage drops in the supply cables within its specified limits.
Particularly with longer cables and higher load currents, this can maintain a considerably more stable supply voltage at the load.
However, correct technical design is essential.
Remote Sense does not eliminate the electrical resistance of connecting cables or the power losses caused by load current.
The cables must therefore still be adequately sized and electrically protected.
The additional output voltage and power must also remain within the permissible instrument limits.
Correct assignment of Force and Sense leads and the proper sensing point directly at the intended load are particularly important.
An incorrectly connected or interrupted Sense lead can impair regulation and, under certain conditions, cause dangerous output voltages.
Suitable overvoltage protection and clearly defined fault responses are therefore important selection criteria.
Dynamic behaviour must also be considered.
Long cables, load steps and additional input capacitances can affect control loop stability.
A laboratory power supply with Remote Sense must therefore not only compensate for static voltage drops but also operate stably under the actual load conditions.
Finally, the function must not be confused with ordinary remote control or a monitor output.
The selected instrument must explicitly support suitable remote voltage sensing.
Define load voltage and maximum current → Determine cable length and cross-section → Calculate cable resistance → Determine maximum voltage drop → Select a power supply with genuine Remote Sense functionality → Check compensation and power limits → Define the Sense connection point at the load → Verify polarity and EMC-compliant cable routing → Evaluate overvoltage protection and Sense fault behaviour → Check voltage under load → Investigate dynamic load changes → Document the results
The most important practical principle is therefore: What matters is not only the voltage at the output terminals of the laboratory power supply but the actual supply voltage available at the device under test. Remote Sense can stabilise this voltage within its technical limits, provided that the instrument, cables, protective functions and regulation behaviour are properly designed.
24. Frequently Asked Questions About Remote Sense on Laboratory Power Supplies
24.1 What Does Remote Sense Mean on a Laboratory Power Supply?
Remote Sense refers to the remote voltage sensing function of a suitable power supply. Two additional Sense leads measure the voltage directly at the remote load. The power supply can adjust its output voltage accordingly and compensate for voltage drops in the current-carrying supply cables.
24.2 Why Is Remote Sense Useful with Long Cables?
Longer cables have higher electrical resistance. At a given load current, this causes a greater voltage drop. Remote Sense makes it possible to regulate the voltage at the load rather than exclusively at the output terminals of the power supply.
24.3 How Many Conductors Are Required for Remote Sense?
A typical Remote Sense connection uses four conductors: two current-carrying supply cables and two additional voltage sensing leads. These are frequently designated Output +, Output −, Sense + and Sense −.
24.4 Where Are the Sense Leads Connected?
The Sense leads are connected to the voltage sensing points specified by the manufacturer, normally as close as possible to the supply terminals of the device under test. Sense + connects to the positive potential and Sense − to the negative potential.
24.5 Can Sense Leads Carry the Normal Load Current?
No. In a correctly designed Remote Sense power supply, the Sense leads are used for high-impedance voltage measurement. The actual load current continues to flow through the appropriately rated positive and negative supply conductors.
24.6 Does Remote Sense Eliminate the Resistance of the Supply Cables?
No. Cable resistance remains unchanged. Remote Sense can compensate for the corresponding voltage drop by increasing the output voltage of the power supply. Cable power losses and heat generation also remain.
24.7 How Is the Voltage Drop of a DC Supply Calculated?
Voltage drop is calculated by multiplying load current by total cable resistance: ΔU = I · R. For a two-wire DC supply, both the outgoing and return conductors must be considered, together with any additional contact resistances.
24.8 Can Remote Sense Compensate for Any Voltage Drop?
No. The compensation capability is limited by the specifications of the respective power supply. Particularly important parameters include maximum permissible output voltage, output power, current limit and, where applicable, a separate Remote Sense compensation limit.
24.9 Why Is the Power Supply Output Voltage Higher Than the Setpoint with Remote Sense Enabled?
This is normal within the specified compensation limits. The power supply increases its output voltage by the amount lost across the supply cables. This allows the desired voltage to be maintained at the remote load.
24.10 Does Remote Sense Also Work in Constant Current Mode?
Remote Sense is primarily a function for regulating load voltage in constant voltage mode. Once the current limit is reached and the power supply operates in CC mode, voltage regulation cannot simultaneously force the load voltage to reach the intended CV setpoint. The exact function of the Sense inputs in other operating modes depends on the instrument.
24.11 What Happens If a Sense Lead Is Interrupted?
The behaviour depends on the power supply. Without suitable protective measures, incorrect voltage feedback can cause undesirable output voltages. Some power supplies therefore provide open-sense protection, automatic return to local regulation or corresponding fault shutdown functions.
24.12 Is Overvoltage Protection Important When Using Remote Sense?
Yes. Suitable overvoltage protection is particularly important for sensitive devices under test. However, the OVP setting must account for the fact that the power supply output voltage may be higher than the desired load voltage during normal compensation. The manufacturer’s specifications are decisive.
24.13 Can Remote Sense Cause Oscillations?
Under unfavourable conditions, long cables, additional inductance and capacitance, and dynamic loads can affect control loop stability. The manufacturer’s requirements concerning cable length, routing and permissible load conditions must therefore be observed.
24.14 Which Cables Are Required for the Sense Connection?
The Sense leads must be suitable for the voltage, environment and respective manufacturer instructions. Depending on the instrument, twisted pairs, specific shielding or defined routing together with the supply cables may be required. A large conductor cross-section such as that used for current-carrying power cables is normally unnecessary.
24.15 Can I Supply Multiple Loads with a Remote Sense Power Supply?
Suitable power supplies can generally supply multiple loads. However, a single Remote Sense control loop normally refers to one defined voltage sensing point. Voltage drops to other loads are not automatically compensated individually. A suitable distribution arrangement or multichannel concept may therefore be necessary.
24.16 Is a Remote Control Input the Same as Remote Sense?
No. A remote control input can be used to adjust voltage or current externally. Remote Sense, on the other hand, measures the voltage actually present at the remote load and uses it for automatic voltage regulation. These two functions are electrically and functionally different.
24.17 Which Laboratory Power Supplies Does ICS Schneider Offer for These Applications?
ICS Schneider offers various regulated laboratory power supplies, including the 2257.1, 2258.1 and 2258.7 high-power DC voltage regulators and multichannel supplies such as the 2229.5. Depending on the configuration, these instruments have different voltage, current and power ranges. Genuine Remote Sense functionality is not explicitly confirmed in the published product specifications reviewed for these models and must be clarified separately if required.
24.18 What Information Does ICS Schneider Need to Select a Remote Sense Laboratory Power Supply?
The required information includes the voltage needed directly at the load, normal and maximum load current, possible inrush or peak currents and permissible voltage deviation. The distance between the power supply and the load, conductor cross-section, cable material, existing connectors and switching contacts, and calculated voltage drop are particularly important. Information about the required compensation voltage, output power, CV/CC functions, dynamic load changes and possible input capacitance of the device under test is also needed. For safe selection, the desired Sense lead monitoring, overvoltage protection, EMC conditions, number of loads, any required multiple output channels and requirements concerning measurement accuracy, documentation and interfaces must also be specified. Explicit confirmation that the selected instrument configuration supports genuine Remote Sense voltage regulation is essential.