WIKA Model ALM Level sensor with reed chain technology For Nuclear Power Plants


  • Process- and system-specific solutions possible
  • Operating conditions:
    – Operating temperature: T = -10°C …+70 °C (+158 °F)
    – Operating pressure: ambient
  • Lengths: up to 20 m (65 ft)
  • Humidity: 100 %
  • Operational radiation dose: ≤ 160 kGy (16 MRad)
Datasheet

Applications

  • Power generating equipment, power plants, nuclear power plants
  • Water basins
  • Cooling water tanks/pools

 

Accident conditions

  • Accident temperature: 156 °C (312 °F)
  • Accident pressure: 7 bar (0.7 MPa)
  • Accident mission time: 1 year
  • Accident rad. dose: integrated 5.05 MGy (505 MRad)
  • Seismic acceleration: up to 2.5 g (higher accelerations up to 5 g on demand)

Description

ALM level sensors are used to measure and transmit the level of liquids in conjunction with a customer’s control unit outside of the containment area. It is based on the float principle with magnetic transmission in a 3-wire potentiometer circuit.

The resistance measuring chain is closely stepped. Due to this assembly the generated voltage is approximately continuous. Signal transmission over large distances and use in hazardous areas are possible.









Technical advice: 03303 / 50 40 66

Technical advice for this product


    In accordance with the General Data Protection Regulation (GDPR), we are obliged to inform you about the collection of data. We meet this obligation by referring to our privacy notice.



    WIKA Model ALM Level sensor with reed chain technology For Nuclear Power Plants

    • Process- and system-specific solutions possible
    • Operating conditions:
      – Operating temperature: T = -10°C …+70 °C (+158 °F)
      – Operating pressure: ambient
    • Lengths: up to 20 m (65 ft)
    • Humidity: 100 %
    • Operational radiation dose: ≤ 160 kGy (16 MRad)
    Datasheet

    Applications

    • Power generating equipment, power plants, nuclear power plants
    • Water basins
    • Cooling water tanks/pools

     

    Accident conditions

    • Accident temperature: 156 °C (312 °F)
    • Accident pressure: 7 bar (0.7 MPa)
    • Accident mission time: 1 year
    • Accident rad. dose: integrated 5.05 MGy (505 MRad)
    • Seismic acceleration: up to 2.5 g (higher accelerations up to 5 g on demand)

    Description

    ALM level sensors are used to measure and transmit the level of liquids in conjunction with a customer’s control unit outside of the containment area. It is based on the float principle with magnetic transmission in a 3-wire potentiometer circuit.

    The resistance measuring chain is closely stepped. Due to this assembly the generated voltage is approximately continuous. Signal transmission over large distances and use in hazardous areas are possible.









    Technical advice: 03303 / 50 40 66

    Technical advice for this product


      In accordance with the General Data Protection Regulation (GDPR), we are obliged to inform you about the collection of data. We meet this obligation by referring to our privacy notice.



      Diese Website benutzt Cookies. Wenn du die Website weiter nutzt, gehen wir von deinem Einverständnis aus.