- Single-channel pressure controller housing
- Compatible with any interchangeable PACE CM module
- Ideal for end-of-line tests as well as laboratory and workshop environments due to compact design and high control stability
- Highest precision: Measurement accuracy up to 0.001% of full scale (FS) for the most demanding calibration and testing applications.
- Pressure range: 25 mbar – 210 bar
- Industry-leading controller performance (speed, accuracy, and stability) for both small and large volumes
- Highest measurement accuracy with minimal measurement uncertainty (12-month specification)
- Completely redesigned user interface with larger display area
- Simple and intuitive menu structure with context-sensitive help
- Test bench connectivity with HDMI monitor output, mouse and keyboard
- “Drop-in” replacement for PACE5000 with backward compatibility of control modules
- Improved diagnostic tools
- Windows USB driver with automatic installation
- Standards-compliant communication: GPIB IEEE-488, RS232, SCPI-99, USBTMC, LXI 1.6 standard (VXI-11 & HiSLIP), NI IVI-C and LabVIEW-certified instrument drivers
Datasheet |
User Manual |
- Dual-channel pressure controller housing
- Can operate with two CM modules in single, auto-ranging, or dual mode
- No restriction on pressure ratio between channels
- High flexibility – suitable for both industrial and laboratory use
- Highest precision: Measurement accuracy up to 0.001% of full scale (FS) for the most demanding calibration and testing applications.
- Pressure range: 25 mbar – 210 bar
- Industry-leading controller performance (speed, accuracy, and stability) for both small and large volumes
- Highest measurement accuracy with minimal measurement uncertainty (12-month specification)
- Completely redesigned user interface with larger display area
- Simple and intuitive menu structure with context-sensitive help
- Test bench connectivity with HDMI monitor output, mouse and keyboard
- “Drop-in” replacement for PACE5000 with backward compatibility of control modules
- Improved diagnostic tools
- Windows USB driver with automatic installation
- Standards-compliant communication: GPIB IEEE-488, RS232, SCPI-99, USBTMC, LXI 1.6 standard (VXI-11 & HiSLIP), NI IVI-C and LabVIEW-certified instrument drivers
Datasheet |
User Manual |
- Total accuracy of 3 Pa / 0.03 mbar using Druck TERPS resonant silicon technology
- Configurable for a wide range of applications with 1, 2, or 3 reference sensors from 200 kPa to 21.1 MPa
- Comprehensive yet easy-to-use platform—from laboratory barometer to practical calibration standard
- Compact, robust transfer standard ideal for on-site testing and interlaboratory comparisons
- Versatile performance maintained over a wide range of environmental conditions and test media
- Precision of 10 ppm Full Scale (FS)
- Long-term stability up to 15 ppm Full Scale (FS) per year
- Optional barometric reference available
Datasheet |
User Manual |
- Choice of enclosures and interchangeable control modules
- Configurations with single, dual, or auto-range control modules
- High-speed pressure control – 5 seconds to fill 300 CC*
- Up to 210 bar (3000 psi / 21 MPa) gauge and absolute pressure
- Precision up to 0.001 % of full scale (FS) over the calibrated temperature range
- Accuracy – see CM3 accuracy table:
- 2, 3.5 bar: 0.0016 % of reading + 0.0033 % FS
- 8–71 bar: 0.0011 % of reading + 0.0026 % FS
- Long-term stability from 0.001 % of full scale (FS)
- Optional barometric reference
- Utilizes DRUCK’s new, unique sensor range
- 28 selectable pressure units and four user-defined units
- Switch test, leak test, test program, burst test, analog output, and voltage-free (dry contact) relays
- Aeronautical option
- Negative calibration of the instrument included as standard
- High-resolution color touchscreen for operation
- Intuitive, task-oriented menu structure
- Compatible with software packages
- RS232, IEEE connectivity, Ethernet, and USB as standard
Datasheet |
User Manual |
- Dual-channel pressure controller housing (for use with two PACE CM control modules)
- Supports single, dual, and auto-range pressure control
- High pressure control speed – fills 300 cm³ in just 5 seconds*
- Pressure ranges up to 210 bar (3000 psi / 21 MPa), gauge and absolute
- Precision up to 0.001% of full scale (FS) over the calibrated temperature range
- Long-term stability from 0.001% of full scale (FS)
- Optional barometric reference available
- 28 selectable pressure units and four user-defined units
- Automated functions: switch test, leak test, test program, burst test
- Analog output signals and voltage-free (dry contact) relay outputs
- Aeronautical option for pressure, altitude, airspeed, Mach number, and rate of climb (PACE 6000 only)
- Negative calibration of the measurement range included as standard
- 7-inch color touchscreen with VGA resolution
- Intuitive, task-based menu navigation
- Compatible with calibration and control software
- RS232, IEEE-488, Ethernet, and USB as standard communication interfaces
- Modular system with interchangeable CM control modules
- Wide selection of pressure ranges and accuracy classes (Standard, High, Premium, Reference)
- Digital control for high stability and fast response times
Datasheet |
User Manual |
Calibration Manual |
- Single or multiple pressure range configurations available
- Up to three pressures displayed simultaneously
- Utilizes Druck’s innovative piezo-resistive and resonant pressure sensor technology
- Pressure measurement ranges up to 1000 bar (14,500 psi / 100 MPa)
- Precision selectable up to 0.001% Full Scale (FS)
- Long-term stability up to 0.001% FS per year
- Optional barometric reference
- Airfield task as standard with barometer option: Display QFE, QFF, or QNH in pressure units or altitude in feet/meters
- Negative gauge calibration included as standard
- Data logging as standard with on-screen replay
- Selectable numeric or graphic display
- High-resolution touchscreen
- Intuitive icon-driven menu structure
- Switch test/analogue output option
- Switch test/voltage-free contact option
- Aeronautical option
- Leak test option
- RS232, IEEE, Ethernet, and USB interfaces as standard
- Min/Max/Average value display
- Compatible with software packages
- 28 selectable pressure units plus 4 user-defined units
- Various service and support options available
Datasheet |
User Manual |
- Pressure ranges: 1 ... 1,000 mbar (0.01 ... 14.5 psi)
- Accuracy: to 0.1 % FS
- Integrated, automatic pressure generation
- Portable, also usable without power supply unit due to Li-Ion battery
- Simple operation
Datasheet |
User Manual |
- Pressure ranges: -1 ... 210 bar (-15 ... 3,045 psi)
- Control speed 10 s
- Control stability < 0.005 % FS
- Accuracy to 0.02 % IS (IntelliScale)
- Precision 0.008 % FS
Datasheet |
User Manual |
- Pressure ranges: 0 ... 700 bar (0 ... 10,000 psi)
- Control speed 30 s
- Control stability 0.008 % FS
- Accuracy down to 0.01 % IS (IntelliScale)
Datasheet |
User Manual |
- Pressure ranges: -1 ... 210 bar (-15 ... 3,045 psi)
- Control speed 15 s
- Control stability < 0.003 % FS (typical 0.001 % FS)
- Accuracy down to 0.01 % IS (IntelliScale)
- Precision 0.004 % FS
Datasheet
|
User Manual
|
- Pressure ranges: -1 ... 400 bar [-15 ... 6,000 psi] in the pressure types positive and negative gauge pressure al well as absolute pressure
- Up to three integrated, interchangeable reference pressure sensors
- Control stability 0.002 % of the span
- Accuracy down to 0.008 % IS (IntelliScale)
- Precision 0.004 % FS
Datasheet |
User Manual |
- Pressure ranges: 75 ... 10,000 psi up to 290 ... 23,000 psi (5 ... 700 bar up to 20 ... 1,600 bar)
- Pressure medium: Hydraulic oil or water
- Control stability: 0.005 % of FS
- Accuracy: to 0.01 % of FS
- Interchangeable reference pressure sensors
Datasheet |
User Manual |
- Pressure ranges: -1 ... 210 bar [-15 ... 3,045 psi]
- Control speed < 4 seconds for 25 % FS steps
- Accuracy: to 0.02 % FS
- PACE emulation
Datasheet
|
User Manual
|
- Automated and self-contained pressure generation and control from -0.9 bar to 70 bar (1,000) psi
- Standard accuracy to 0.02% of Span*
- Optional precision accuracy models up to 0.01% of Span
- Two removable internal pressure modules for multi-range selection
- Control stability to 0.003% of Span
- Portable, designed for use in the field and in the lab
- Ability to measure two external pressure modules
- Wi-Fi, Bluetooth, USB and Ethernet communication
- HART and profibus communication
- Data logging and task management
- Patented electric pump technology and improved speed
Datasheet |
User Manual |
- Automated pressure generation and control to 10,000 psi (700 Bar)
- Accuracy to 0.01%FS
- Dual-range manual or auto select
- Control stability ≤ 0.005%FS
- Portable - designed for use in the field and in the lab
- Control by optional external pressure modules
- Supports two external pressure modules
- Wi-Fi, LAN, Bluetooth, USB and Ethernet communication
- Full HART field communicator
- HART and PROFIBUS communication
- Data logging and task management
- Patented electric pump technology
Datasheet |
- Individuelle Prüfeinrichtung für Werkstatt, Prozess und Service
- Automatische Druckregler PACE 5000/ PACE 6000 integriert
- Für wählbare Messbereiche von -1 … 0 … +210 bar
- Automatische Vorgabe der Prüfpunkte, Rampen über Touchscreen
- Genauigkeit: 0,05 bis zu 0,0015 % FS
- Großes TFT-Touchscreen
- Eingebaute N2-Stickstoffflasche mit präzisem manuellen Druckregler (auf Wunsch externe Druckflasche)
- Eingebaute Vakuumpumpe für Unterdruck / Absolutdruck (auf Wunsch auch ohne Vakuumpumpe)
- Eingebaut analog Manometer für Vakuum, Überdruck und Flaschendruck
- Externe Nachfüllung der N2_Stickstoffflasche über Minimess
- Prüflingsanschluss über Minimess-Schnellkupplung
- Betrieb an Netz 230 V AC 50 Hz oder Batterie-Betrieb
- Robuster abschließbarer Alukoffer
Datenblatt
|
.
.
- Interchangeable reference sensors that can be changed via the front
- Large colour touchscreen with all necessary information
- Control stability 0.001 % of span
- Accuracy up to 0.009 % IS-50 (IntelliScale)
Datasheet |
- On-site selection of pressure range from the GENii PM620/PM620T pressure module family.
- Built-in barometer for accurate pseudo-pressure measurement.
- Hybrid pressure generation and control from -0.9 barg (-13 psi) to 20 barg (300 psi).
- Automatic selection between manual pressure generation, automatic generation, or venting with large test volume capacity.
- Fully automatic pressure generation and control for precise setpoint maintenance.
- Minimization of field leaks using quick-fit pressure adapters.
- Physical or *Bluetooth® wireless connection to a DPI620G electrical calibrator.
- Field-replaceable rechargeable battery pack.
- Autonomous calibration assistant for storing procedures and calibration data.
Datasheet |
User Manual |
- Pressure ranges from -14.35 psi (-0.99 bar) to 3600 psi (250 bar).
- Two removable internal pressure modules with multi-range selection.
- Available accuracies of 0.02% FS, 0.1% FS, and 0.011% RD.
- Control stability of 0.003% FS.
- Ultra-fast pressure control.
- Absolute and gauge pressure measurement.
- WiFi-enabled communication.
- Large 7-inch color touchscreen.
- Optional barometric module.
- User-friendly, icon-driven interface.
- I/O alarm function.
Datasheet |
Datasheet |
.
.
- Pressure ranges from vacuum to 200 bar (3000 psi)
- Removable interchangeable intelligent sensors
- Precision accuracy of 0.005% of reading +0.005% of full scale
- Standalone solution up to 70 bar, no gas bottle required when used with the electric pump
- Standard model accuracy of 0.02% of full scale
- External pressure modules to 200 bar (3000 psi)
- WiFi enabled communications
- Fully temperature compensated accuracy from 0°C to 50°C
- HART communication and Profibus PA
- Large 7” color touch screen display
- Control stability of 0.003% FS
- Built-in barometer
- Easy-to-use icon based user interface
Datasheet |
- Gas Pressure System
- Range from -1 to 1000 bar
- Setting the variable volume with hub
- Pressure Gauge (digital or analog possible)
Automatic Pressure Regulators – precise, fast & repeatable pressure control
Automatic pressure regulators (electronic pressure regulator / pressure controller) close the loop to control pressure automatically, executing setpoints and profiles without manual intervention. They deliver stable, repeatable results, cut operator error, and speed up calibration, leak and functional tests in laboratory, production and service.
Key capabilities (model-dependent): ranges from vacuum to 400 bar pneumatically (higher hydraulically), accuracies to ±0.01 % FS/Reading, step/ramp/leak/sweep profiles, dual-range/auto-range, integrated reference sensors or remote modules, media paths for gas/oil/water, interfaces SCPI/Modbus/TCP, analog I/O, HART passthrough, plus data logging and CMMS/CAQ connectivity.
ICS Schneider Messtechnik supplies turnkey solutions including adapter/hose kits, manifolds, safety concepts, software and traceable calibration (factory/DAkkS).
FAQ on Automatic Pressure Regulators
Answers on control behavior, accuracy, media, connections, safety, software and best practices.
How do automatic regulators differ from simple pressure reducers?
Automatic regulators act actively with valves/sensors in a closed control loop. They hold the setpoint even with leaks or volume changes and can run sequences and profiles.
Which control profiles are common?
| Profile | Description | Use case |
|---|---|---|
| Step | Discrete levels with dwell time | Calibration points (0–25–50–75–100 %) |
| Ramp | Linear rise/fall | Leak/functional testing |
| Leak/Hold | Hold setpoint & measure drift | Leak tests (mbar/min) |
| Sweep | Cyclic up/down runs | Hysteresis/repeatability |
What accuracy can I expect?
Depending on model ±0.01…0.05 % FS/Reading. For pass/fail decisions target a TUR ≥ 4:1 (e.g., 0.25 % DUT → ≤0.06 % reference).
How do I size the range properly?
Choose a span so your working range uses 60–90 % of the regulator’s span. For wide coverage use dual-range or exchangeable sensor modules.
Gas or hydraulic—what fits my application?
| Medium | Advantages | Notes |
|---|---|---|
| Gas (air/N₂) | Fast, clean, fine control | Manage venting/emissions; limited max pressure |
| Hydraulic (oil/water) | High pressures, low compressibility | Ensure cleanliness, compatibility and proper bleeding |
What connections are typical?
G 1/4 (BSPP), NPT, M20×1.5 and sometimes 7/16″-20 UNF. Seal via O-ring/face seal (BSPP) or taper/cone (NPT/UNF). Avoid tape on sealing faces.
Do regulators handle gauge, absolute and differential?
Many support gauge/absolute (with a barometric reference). Δp requires a suitable manifold with equalize/isolate.
How stable is the held pressure?
Gas setups reach Pa–mbar stability; hydraulic systems typically hold within the mbar range. Minimize system volume, temperature drift and vibration.
Can I integrate the regulator into automated rigs?
Yes—via SCPI/Modbus/TCP, drivers for LabVIEW/.NET or REST APIs. Integrate sequences, data capture and barcode/asset IDs; export CSV/PDF, maintain an audit trail and roles; optional 21 CFR Part 11 e-signatures.
What does a typical procedure look like?
- Connect the DUT, select medium, bleed/degas.
- Run a sequence (e.g., 0–25–50–75–100 % up/down).
- Regulator holds the setpoint; after the dwell time capture results.
- Adjust the DUT if needed; repeat and document as-left.
How are leak tests performed?
Drive the setpoint, switch to hold, measure drift rate (e.g., mbar/min) over a defined interval; log temperature and compensate if required.
Maintenance & calibration—what should I plan for?
Regular factory/DAkkS calibration, check filters/seals, keep hydraulic media clean. Keep firmware/procedures up to date.
Typical issues & remedies
- Slow regulation → reduce hose volume, optimize control band/valve timing.
- Offset → verify units/reference (abs/g), enable barometric correction.
- Leaks → clean sealing faces, replace O-rings, torque fittings correctly.
Do you provide accessories & documentation?
Yes: manifolds/adapters, hose kits, carry cases, material/media approvals, calibration certificates (factory/DAkkS) and IQ/OQ checklists.
Do you assist with selection, integration & training?
Yes—we size ranges/media, plan interfaces, integrate software/CMMS, prepare procedures and train your team—turnkey.












































































































































































Datasheet
Datasheet