PPS1210 – Druckkalibrator


  • Pressure range: 0 – 700 bar
  • Max. Permitted pressure: FS + 10%
  • Resolution: 100 mBar (10 mBar at pressure range = 200 bar)
  • Accuracy: ± 0.05% f.s. E. (at t = 0 °C to +50 °C) +/- 1 digit
  • Measuring rate: 500 ms (2 Hz)
  • Units: including Bar / PSI / kPA / mH2O (a total of 13 different units, adjustable on the display)
  • Heated pressure port : 1/2″ 20UNF or M18 x 1.5 (other ports on request)
  • Temperature at the pressure port: Adjustable from RT to +400°C on the built-in PID temperature controller
  • Oil volume: 57 ml
  • Type of oil: olive oil/HLP 22 BP
  • Power supply: 3 V battery, type CR 2430
  • Battery life: 2000 hours of continuous use
  • Interface: RS485, connecting cable K-114-A with USB port optionally available
  • Power supply for heating and sensor supply: 115 V AC or 230 V AC
  • Sensor supply (test object): via integrated power pack and plug or socket
  • Protection class: IP 65
Datasheet
User Manual
Technical advice: 03303 / 50 40 66

Technical advice for this product


    In accordance with the General Data Protection Regulation (GDPR), we are obliged to inform you about the collection of data. We meet this obligation by referring to our privacy notice.



    PPS1210 – Druckkalibrator

    • Pressure range: 0 – 700 bar
    • Max. Permitted pressure: FS + 10%
    • Resolution: 100 mBar (10 mBar at pressure range = 200 bar)
    • Accuracy: ± 0.05% f.s. E. (at t = 0 °C to +50 °C) +/- 1 digit
    • Measuring rate: 500 ms (2 Hz)
    • Units: including Bar / PSI / kPA / mH2O (a total of 13 different units, adjustable on the display)
    • Heated pressure port : 1/2″ 20UNF or M18 x 1.5 (other ports on request)
    • Temperature at the pressure port: Adjustable from RT to +400°C on the built-in PID temperature controller
    • Oil volume: 57 ml
    • Type of oil: olive oil/HLP 22 BP
    • Power supply: 3 V battery, type CR 2430
    • Battery life: 2000 hours of continuous use
    • Interface: RS485, connecting cable K-114-A with USB port optionally available
    • Power supply for heating and sensor supply: 115 V AC or 230 V AC
    • Sensor supply (test object): via integrated power pack and plug or socket
    • Protection class: IP 65