The DPI 705 combines outstanding portability with exceptional robustness.
Engineered to fit comfortably in the technician’s hand, this precision instrument is designed for demanding industrial use.
Its light weight, compact design, and intuitive operation make it your constant companion.
At the same time, its splash-proof ABS housing (IP54) provides excellent durability.
The DPI 705 delivers high-accuracy measurements across a wide pressure range.
The large, easy-to-read LCD display shows a 5-digit measurement value.
With a high accuracy of 0.1%, the DPI 705 serves as a reliable pressure reference instrument.
When used with optional pneumatic or hydraulic hand pump connection kits, the unit can be extended into a full-featured calibrator.
As a universal instrument, the DPI 705 is available in ranges from 0–70 mbar up to 0–700 bar.
Versions are available for gauge, absolute, and differential pressure measurement, both with internal and external sensors.
The stainless-steel sensor, manufactured in-house, has proven its reliability in millions of tough industrial applications.
The DPI 705 supports your applications with convenient functions.
The following features enhance flexibility for daily measurement tasks in plant engineering, service, production, and development:
Selectable at the touch of a button:
16 pressure units: mbar, bar, psi, Pa, kPa, MPa, kg/cm², in Hg, mmHg, mH₂O, inH₂O, ftH₂O (both at 4° and 20 °C)
Intrinsic Safety Version
With the intrinsically safe version DPI 705/IS, new possibilities open up for the user.
Lengthy applications for Ex entry permits are no longer required — with the DPI 705/IS you can safely operate even in Ex Zone 0.
Pressure Ranges
The version with an internal pressure sensor features a G⅛ female pressure connection; a 6 mm hose adapter is pre-installed.
The version with an external pressure sensor uses an electrical connector with a 1 m cable.
All sensors have a G¼ female pressure connection.
| Pressure Range | Internal Sensor | External Sensor |
|---|---|---|
| 0–70, 200 mbar | Differential | Gauge / Absolute / Differential |
| 0–350 mbar | Gauge | Gauge / Absolute / Differential |
| 0–700 mbar | Differential | Gauge / Absolute / Differential |
| 0–1 bar | Gauge / Absolute | Gauge / Absolute / Differential |
| 0–2 bar | — | Gauge / Absolute / Differential |
| 0–3.5 bar | — | Gauge / Absolute / Differential |
| 0–7 bar | Gauge | Gauge / Absolute / Differential |
| 0–10 bar | Gauge | Gauge / Absolute / Differential |
| 0–20 bar | — | Gauge / Absolute / Differential |
| 0–35 bar | — | Gauge / Absolute / Differential |
| 0–70 bar | — | Gauge / Absolute |
| 0–135, 200, 350, 700 bar | — | Absolute |
Technical Data (excerpt)
Static Pressure for Differential Ranges
- Internal sensor: max. 2 bar (IS version: max. 1.1 bar)
- External sensor: max. 35 bar
Accuracy
- Non-linearity, hysteresis: ±0.1% of range
- Long-term stability: ±0.1% of range per year
Overpressure
- 2× overload without permanent calibration shift
- Display flashes at 110% of range
Temperature Effect
- 0 to 50 °C: ±0.02% of reading per °C
- Additionally, for ranges > 1 bar: ±0.02% of span per °C
- For 1 bar absolute: ±0.05% of span per °C
Pressure Media
- Gauge, Absolute: Gases and liquids compatible with stainless steel 1.4401
- Differential (internal): Dry, non-corrosive gases
- Differential (external positive port): Gases and liquids compatible with stainless steel 1.4401
Power Supply
- 3 × 1.5 V AA alkaline cells, operating time > 300 hours
Electrical Safety / EMC
- Protection class IP54, EN 61010, EN 50081-1, EN 50082-2
Intrinsically Safe Version
- EEx ia II C T4 (−10 to +50 °C)
- According to EN 50020:1994 and EN 50014:1992
- With CAN/CSA E79-11-95 and CAN/CSA E79-0-95 (Division 1, Class 1, Groups A, B, C and D)
Options
Pneumatic / Hydraulic Connection Kits
Supplied in a hard carrying case with foam insert.
Pneumatic kits include the PV111 hand pump, range −0.8 to +25 bar,
with pressure hose and adapter set.
Hydraulic kits include the PV112 hand pump, range 0 to 700 bar,
with pressure hose and adapter set.
- (A1) Pneumatic kit with adapters G⅛, ⅜, ¼, ½
- (A2) Hydraulic kit with adapters G⅛, ⅜, ¼, ½
- (A3) Pneumatic kit with adapters ⅛, ¼, ½ NPT
- (A4) Hydraulic kit with adapters ⅛, ¼, ½ NPT
(B) Carry Pouch
Practical protective pouch for the DPI 705 with belt loop and accessory pocket.
Also accommodates an external pressure sensor.
(D) Additional Vacuum Calibration
- Ranges < 1 bar are calibrated symmetrically (e.g. ±200 mbar).
- Ranges ≥ 1 bar are calibrated down to −1 bar (e.g. −1 to +20 bar).
Ordering Information
- Select standard (DPI 705) or intrinsically safe (DPI 705/IS) version.
- Select internal or external pressure range according to the table.
Note: the instrument supports only one range. - Select gauge, absolute, or differential pressure type.
- Add desired options (A1)–(A4), (B), (D).
Scope of Delivery
The DPI 705 is supplied complete with an operating manual and calibration certificate.
Traceability
All calibration data are recorded using instruments maintained under an ISO 9001 quality system and are therefore traceable to national standards.
Frequently Asked Questions (FAQ) – DPI 705 / DPI 705 IS
Is this product still available?
No – the DPI 705 and DPI 705 IS are no longer offered. Recommended alternative:
DPI 705E Precision Pressure / Temperature Indicator.
What pressure ranges are available?
The instrument covers ranges from 0–70 mbar to 0–700 bar — internally or externally, and in versions for gauge, absolute, and differential pressure.
What is the accuracy?
Accuracy: ±0.1% of full scale (non-linearity + hysteresis). Long-term stability: ±0.1% FS per year.
What features does it offer?
Large, easy-to-read 5-digit LCD display.
Selectable 16 pressure units (mbar, bar, psi, Pa, kPa, MPa, kg/cm², inHg, mmHg, mH₂O, inH₂O, ftH₂O at 4 °C / 20 °C).
Rugged ABS housing, IP54.
Battery-powered (3 × 1.5 V AA) with over 300 hours operation.
Optional: pneumatic/hydraulic connection kits for calibration use.
What is the difference between DPI 705 and DPI 705 IS?
The IS version is intrinsically safe and approved for Ex Zone 0.
The standard version is not intrinsically safe.
What media and connections are supported?
Gauge/Absolute: gases or liquids compatible with stainless steel 1.4401.
Differential (internal): dry, non-corrosive gases only.
Internal connection: G⅛ female (adapter for 6 mm hose).
External connection: G¼ female with 1 m cable.
How resistant is the device to overpressure?
Overpressure protection up to 2× full scale without permanent change.
Display flashes at 110% FS.
How does temperature affect the measurement?
Temperature coefficient:
±0.02% of reading per °C (0 … 50 °C).
For ranges > 1 bar: ±0.02% of span per °C.
For 1 bar absolute: ±0.05% of span per °C.
What protection class and standards apply?
Protection class IP54.
Electrical safety / EMC in accordance with EN 61010, EN 50081-1, EN 50082-2.
DPI 705 IS: EEx ia II C T4 (−10 … +50 °C) per EN 50020 / EN 50014, CAN/CSA E79-11-95, E79-0-95.
What options and accessories are available?
Pneumatic connection kits with PV111 hand pump (−0.8 … +25 bar).
Hydraulic connection kits with PV112 hand pump (0 … 700 bar).
Protective carry pouch with accessory pocket (Code B).
Vacuum calibration option (Code D).
What is included in delivery?
Instrument, operating manual, and calibration certificate.
Calibration data are traceable to national standards (ISO 9001).
How do I select the correct version?
1. Choose standard (DPI 705) or intrinsically safe (DPI 705 IS).
2. Define internal or external range.
3. Specify pressure type (gauge, absolute, or differential).
4. Add option codes (A1–A4, B, D).
Are there any limitations or notes?
Product no longer available; DPI 705E recommended.
Static pressure: internal max. 2 bar (IS 1.1 bar), external max. 35 bar.
Differential pressure sensor: dry, non-corrosive gases only.
Measurement range is fixed and non-adjustable.
What are the main application areas?
Designed for industrial measurement, service, and maintenance applications.
Its rugged design and battery operation make it ideal for field use.












































































































































































Datasheet