DPI610E-Aero is a cost-effective yet versatile portable calibrator for precise leak testing of aircraft Pitot-static systems.
Configured as a single-channel instrument – selectable for either altitude or airspeed testing – it is fully self-contained with manual pressure and vacuum generation as well as venting capability. The conversion to aviation-specific units is integrated; the instrument also provides leak and switch tests along with additional functions.
The pneumatic system of the DPI610E-A limits the maximum rate of climb and descent to protect sensitive aircraft instruments such as rate-of-climb/vertical speed indicators (RoC/VSI).
With a 5 m aviation hose (6 mm internal diameter) and a connected RoC instrument, and a total volume of 1 litre, the maximum rate of climb is ± 6,000 ft/min (the same applies to the descent rate).
The pressure connector can be supplied with AN4, AN6, Hansen or Stäubli quick couplings. An integrated valve allows controlled pressure venting to ambient pressure on the ground.
Standard specifications
Aviation ranges
- Altitude: −3,000 to 50,000 ft
- Airspeed: 0 to 600 kn
Aviation scales
- Altitude: feet and metres (or pressure units)
- Airspeed: knots, km/h, or pressure units (mbar, hPa, kPa, psi, mmHg@0 °C, inHg@0 °C, inH₂O@0 °C)
Leak test scales
- Airspeed: kn/min, km/h/min, or pressure units per minute (mbar, hPa, kPa, psi, mmHg@0 °C, inHg@0 °C, inH₂O@0 °C)
- Altitude: ft/min, m/min, or pressure units per minute (mbar, hPa, kPa, psi, mmHg@0 °C, inHg@0 °C, inH₂O@0 °C)
Pressure/vacuum source
Integrated pneumatic hand pump and fine adjuster for generating and precisely setting pressure or vacuum.
Relief valve and ground/atmosphere valve for venting to ambient pressure.
The rate of climb/descent is limited to ± 6,000 ft/min when an instrument (including hose) with a total volume of 1 litre (1000 cm³) is connected.
Temperature range
- Operating: −10 to +50 °C
Accuracy – Airspeed
Valid over 0–600 kn:
- 200 kn: ± 0.6 kn
- 400 kn: ± 0.3 kn
- 600 kn: ± 0.15 kn
Accuracy – Altitude
Valid over −3,000 to 50,000 ft:
- Sea level: ± 1 ft
- 10,000 ft: ± 15 ft
- 30,000 ft: ± 30 ft
Accuracy – Absolute range
- 0.02% FS
The stated accuracies include linearity, hysteresis, repeatability, and 12 months of measurement stability over the full operating temperature range.











































































































































































Datasheet