ICS-02S – Compact Temperature Calibrator & Simulator for Field & Lab
The ICS-02S combines measurement and simulation of RTD, thermocouple, and mV/V signals in a handy instrument.
With a basic accuracy of ±0.05%, a backlit 5½-digit display, and an integrated reference junction, it is ideal for
commissioning, maintenance, and on-site calibration work. Four safety sockets (one channel) plus practical accessories
(4 mm safety leads, crocodile clips, TC reference junction plug) come as standard. Power is supplied by 4 × AAA cells,
providing up to 25 h of operation with auto power-off. Housing and scope of delivery are geared towards mobile use.
At a glance:
- RTD (Pt100/200/500/1000, Cu50): Source & measure −200…+800 °C (0.1 °C resolution), 2/3/4-wire measurement, resistance up to 5.5 kΩ with up to 0.01 Ω resolution.
- Thermocouples R, S, K, E, J, T, N, B, A, C, D: source/measure in °C/°F or as mV (source −100…1100 mV, measure 50…550 mV). Internal reference junction ±0.5 °C.
- Voltage: source/measure −10 mV…1.1 V or −35…+35 V DC (±0.05%). Audible continuity test incl. “open” threshold 0.5 kΩ.
- Ergonomics: large, high-contrast display and simple keypad navigation in a robust ABS housing (210 × 100 × 40 mm, approx. 550 g).
Typical applications:
- Calibration and service of transmitters
- Controllers and indicators, field troubleshooting
- Temperature simulation during commissioning
- Training and QA test benches
| Type | ICS-02S |
|---|---|
| Resistance Thermometer (RTD) | |
| Generate & measure RTD signals | Pt100, Pt200, Pt500, Pt1000, Cu50 (−200…0…+800 °C); 0.1 °C |
| Resistance / Source / Measure | 0…4000 Ω or 0…5500 Ω; 0.1 Ω or 0.01 Ω |
| Accuracy (rdg + const.) | ±0.05% |
| Selectable temperature unit | °C / °F |
| Multi-wire measurement | 2 / 3 / 4 |
| Thermocouples (TC) | |
| Generate & measure TC signals | R, S, K, E, J, T, N, B, A, C, D |
| mV source / mV measure | −100…1100 mV / 50…550 mV |
| Also source/measure in °C | Yes |
| Accuracy (rdg + const.) | ±0.05% |
| Selectable temperature unit | °C / °F |
| Internal reference junction | ±0.5 °C |
| Voltage (V) | |
| Generate & measure voltage | −10 mV…1.1 V or −35…+35 V DC |
| Accuracy (rdg + const.) | ±0.05% |
| Continuity (O/C) | O / C |
| Audible continuity | O / C |
| “Open” threshold | 0.5 kΩ |
| Features | |
| Sensor connector | 4 safety sockets (1 channel) incl. 2 sets of 4 mm safety test leads and crocodile clips; TC reference junction plug |
| Power supply | 4 × 1.5 V AAA batteries; operating time approx. 25 h; with auto power-off |
| User interface | Keypad |
| Display | Multi-function LCD, 5½-digit, backlit |
| Housing | |
| Design / scope of delivery | ABS housing with protective padding; set of safety leads; crocodile clips; TC reference junction; manual; test certificate |
| Dimensions (H × W × D) | 210 × 100 × 40 mm |
| Weight | approx. 550 g |











































































































































































Datasheet