OIL CHECK 500


The monitoring system for high-precision measuring of the Vaporous Residual oil content of Compressed air and Nitrogen.

Datasheet
User Manual
Property Value
Measuring Medium Compressed air, nitrogen, (free from aggressive, corrosive, caustic, toxic, flammable, and fire-promoting components). Other gases on request
Measurement Size Residual oil content in mg oil/Norm m³ based on 1.0 bar [abs], +20° C, 0% relative humidity, according to ISO 8573-1
Detectable Substances Hydrocarbons, functional hydrocarbons, aromatics
Application Area After activated carbon filter, after activated carbon adsorber, after oil-free compressing compressor, each with pre-filtration and drying
Ambient Temperature +20 °C… +45 °C, rel. Humidity <= 75% without condensation
Compressed Air Temperature +20 °C… +50 °C
Operating Overpressure 3…9 bar, optional pressure reducer switchable up to 300 bar
Measuring Gas Humidity <= 40% rel. Humidity, pressure dew point max. +10 °C, non-condensable humidity
Compressed Air Connection G 1/4″ internal thread according to ISO 228-1
Measured Values mg/Norm m³, pressure and temperature compensated residual oil vapor content
Measurement Range 0.001…5 mg/m³ (higher measurement ranges on request)
Detection Limit (Residual Oil) 0.001 mg/m³
Measuring Gas Flow approx. 0.5 Norm liter/minute, based on 1.0 bar [abs] and +20 °C, in a relaxed state
Power Supply Plug-in Power Supply 100…240 VAC / 1 Ph. / PE / 50…60 Hz / ± 10%
Outputs Digital Output: RS 485 interface (Modbus-RTU), Ethernet via DS 400 / 500
Analog Output: 4…20 mA (galvanically isolated)
Optional: 2x 4…20 mA analog output (galvanically isolated)
2x alarm output for external alarm column, alarm values freely adjustable
Operating Hours Counter integrated
Dimensions (mm) 200 x 130 x 120 (W x H x D)
Weight approx. 7 kg

 

“Forced Pressure Variation” for Long-term stable measurement results – auto-calibration:

Using the innovative “Forced Pressure Variation” measuring method, the OIL CHECK 500 generates internal reference gas in different mass concentrations. This process can be used to compensate for ageing or contamination-related components in the measurement signal, in particular long-term drift. No wearing parts such as activated carbon filters are required for the generation of zero air. The result is Low maintenance and Long-term stable Measuring.

Service friendly, no downtime:

The sensor unit can be replaced by the customer on site. This eliminates the need to return the entire device for recalibration.

 

Process reliability:

All functions / components are monitored internally. A complete function test report can be printed out using the service software.

On-site calibration:

Calibration can be carried out in the field using test gas cylinders. The service software can be used to generate a
verification report (as-found data) and a calibration report (as-left data) can be generated.

Ideal for Portable Measuring:

Compact Measuring devices, easy sampling and quickly ready for Measuring

Simple installation

Air inlet / air outlet – sampling via PTFE hose
or stainless steel line.
Digital and analogue interfaces as standard.
Connection via M12 plug

Service friendly – no downtime

Loosen four screws – replace sensor unit. At the touch of a button
a complete function test is carried out. This means that
Measuring can be continued almost without interruption.









Technical advice: 03303 / 50 40 66

Technical advice for this product


    In accordance with the General Data Protection Regulation (GDPR), we are obliged to inform you about the collection of data. We meet this obligation by referring to our privacy notice.



    OIL CHECK 500

    The monitoring system for high-precision measuring of the Vaporous Residual oil content of Compressed air and Nitrogen.

    Datasheet
    User Manual
    Property Value
    Measuring Medium Compressed air, nitrogen, (free from aggressive, corrosive, caustic, toxic, flammable, and fire-promoting components). Other gases on request
    Measurement Size Residual oil content in mg oil/Norm m³ based on 1.0 bar [abs], +20° C, 0% relative humidity, according to ISO 8573-1
    Detectable Substances Hydrocarbons, functional hydrocarbons, aromatics
    Application Area After activated carbon filter, after activated carbon adsorber, after oil-free compressing compressor, each with pre-filtration and drying
    Ambient Temperature +20 °C… +45 °C, rel. Humidity <= 75% without condensation
    Compressed Air Temperature +20 °C… +50 °C
    Operating Overpressure 3…9 bar, optional pressure reducer switchable up to 300 bar
    Measuring Gas Humidity <= 40% rel. Humidity, pressure dew point max. +10 °C, non-condensable humidity
    Compressed Air Connection G 1/4″ internal thread according to ISO 228-1
    Measured Values mg/Norm m³, pressure and temperature compensated residual oil vapor content
    Measurement Range 0.001…5 mg/m³ (higher measurement ranges on request)
    Detection Limit (Residual Oil) 0.001 mg/m³
    Measuring Gas Flow approx. 0.5 Norm liter/minute, based on 1.0 bar [abs] and +20 °C, in a relaxed state
    Power Supply Plug-in Power Supply 100…240 VAC / 1 Ph. / PE / 50…60 Hz / ± 10%
    Outputs Digital Output: RS 485 interface (Modbus-RTU), Ethernet via DS 400 / 500
    Analog Output: 4…20 mA (galvanically isolated)
    Optional: 2x 4…20 mA analog output (galvanically isolated)
    2x alarm output for external alarm column, alarm values freely adjustable
    Operating Hours Counter integrated
    Dimensions (mm) 200 x 130 x 120 (W x H x D)
    Weight approx. 7 kg

     

    “Forced Pressure Variation” for Long-term stable measurement results – auto-calibration:

    Using the innovative “Forced Pressure Variation” measuring method, the OIL CHECK 500 generates internal reference gas in different mass concentrations. This process can be used to compensate for ageing or contamination-related components in the measurement signal, in particular long-term drift. No wearing parts such as activated carbon filters are required for the generation of zero air. The result is Low maintenance and Long-term stable Measuring.

    Service friendly, no downtime:

    The sensor unit can be replaced by the customer on site. This eliminates the need to return the entire device for recalibration.

     

    Process reliability:

    All functions / components are monitored internally. A complete function test report can be printed out using the service software.

    On-site calibration:

    Calibration can be carried out in the field using test gas cylinders. The service software can be used to generate a
    verification report (as-found data) and a calibration report (as-left data) can be generated.

    Ideal for Portable Measuring:

    Compact Measuring devices, easy sampling and quickly ready for Measuring

    Simple installation

    Air inlet / air outlet – sampling via PTFE hose
    or stainless steel line.
    Digital and analogue interfaces as standard.
    Connection via M12 plug

    Service friendly – no downtime

    Loosen four screws – replace sensor unit. At the touch of a button
    a complete function test is carried out. This means that
    Measuring can be continued almost without interruption.









    Technical advice: 03303 / 50 40 66

    Technical advice for this product


      In accordance with the General Data Protection Regulation (GDPR), we are obliged to inform you about the collection of data. We meet this obligation by referring to our privacy notice.



      Diese Website benutzt Cookies. Wenn du die Website weiter nutzt, gehen wir von deinem Einverständnis aus.