The IPD40/41 differential pressure transmitters are characterized by a robust design and a significantly high overload capability. The stainless steel membrane and process connection are laser-welded, resulting in high media resistance and long-term stability. Each measuring range is temperature compensated, which in combination with the piezoresistive measuring principle results in a very accurate measurement even at lowest pressures. Various electrical connection options allow easy integration and are protected against dust and water with IP 67.
- Gaseous media
- Liquid media
- Hygiene applications
- Viscous and pasty media
- Aggressive media
- Pressure superimposed level measurement
| Measuring ranges | Smallest measuring range: 0…50 mbar |
| Largest measuring range: 0…25 bar | |
| Negative pressure measuring range: up to -25 bar | |
| Analog output | 4…20 mA, 2-wires |
| 0…10 V, 3-wires | |
| Power supply | 20 mA output: 9…30 V DC |
| 10 V output: 15…30 V DC | |
| Accuracy | ≤ ± 0,2 % FS @ 25 °C |
| Long term stability | ≤ ± 0,1 % FS / Year |
| Static pressure influence | ≤ ± 0,1 % FS/ 10 bar |
| Hysteresis | ≤ ± 0,05 % FS |
| Response time | 200 ms – others on request |
| Swich-on time | < 1 s |
| Temperature coefficient | Zero: ≤ ± 0,02 % FS / Kelvin |
| Span: ≤ ± 0,02 % FS / Kelvin | |
| Medium temperature | -25…120 °C |
| Surrounding temperature | -25…85 °C |
| Storage temperature | -40…85 °C |
| Short-circuit resistance | Permanent |
| Reverse polarity protection | Protection against reverse polarity, but no function |
| Electromagnetic compatibility | Interference emissions and immunity acc. to EN 61326 |
| Wetted materials | Process connection: Stainless steel 1.4404 |
| Sensor: Stainless steel 1.4404 with silicone oil filling | |
| Protection type | IP 67 |
| Weight | Approx. 650 g |












































































































































































Datasheet