Series 6 – Oxymat6


  • Paramagnetic alternating pressure method: Small measuring ranges (0 to 0.5% or 99.5 to 100% O2), absolute linearity
  • Detector element has no contact with the measuring gas: can be used under “harsh conditions,” long service life
  • Physically suppressed zero point by suitable reference gas selection (air or O2), such as 98 to 100% O2 for purity monitoring / air separators
  • Open interface architecture (RS 485, RS 232, PROFIBUS)
  • SIPROM GA network for maintenance and service information (optional)
Datasheet

Specialist for oxygen

The OXYMAT 6 also meets the high standards in demanding applications through modern electronics, ease of operation, and a physical element that is adapted to the measurement task.

Even for the smallest measuring ranges

The SIL 2 classification obtained by OXYMAT 6 makes it easier for plant operators to comply with the often required proof of risk reduction. The OXYMAT 6 measures oxygen using the paramagnetic alternating pressure method. This guarantees perfect linearity and enables parameterization of smallest measuring ranges from 0 to 0.5% (detection limit 50 ppm) up to 0 to 100%, and even 99.5 to 100% in a single device. High-quality materials in the gas channel even enable the analytical devices to be used for measurements of corrosive gas mixtures. The detector unit has no contact with the measuring gas, thus allowing it to be used under harsh conditions while guaranteeing a long service life at the same time.

 

Long lifetime

Gas-tight separation of the electronic and physical components when using the field analyzer design with class IP65 protection and a purging gas connection guarantees a long lifetime even under harsh ambient conditions. Moreover, the field device design is also the basic variant for the optional heating feature of the analysis unit for measuring any low-boiling components and for use in hazardous zones

Technical specifications

Max. number of components
1
Components
Oxygen
Minimal measuring range
0-0,5 %
Housing
19″ rack / field housing








Technical advice: 03303 / 50 40 66

Technical advice for this product


    In accordance with the General Data Protection Regulation (GDPR), we are obliged to inform you about the collection of data. We meet this obligation by referring to our privacy notice.



    Series 6 – Oxymat6

    • Paramagnetic alternating pressure method: Small measuring ranges (0 to 0.5% or 99.5 to 100% O2), absolute linearity
    • Detector element has no contact with the measuring gas: can be used under “harsh conditions,” long service life
    • Physically suppressed zero point by suitable reference gas selection (air or O2), such as 98 to 100% O2 for purity monitoring / air separators
    • Open interface architecture (RS 485, RS 232, PROFIBUS)
    • SIPROM GA network for maintenance and service information (optional)
    Datasheet

    Specialist for oxygen

    The OXYMAT 6 also meets the high standards in demanding applications through modern electronics, ease of operation, and a physical element that is adapted to the measurement task.

    Even for the smallest measuring ranges

    The SIL 2 classification obtained by OXYMAT 6 makes it easier for plant operators to comply with the often required proof of risk reduction. The OXYMAT 6 measures oxygen using the paramagnetic alternating pressure method. This guarantees perfect linearity and enables parameterization of smallest measuring ranges from 0 to 0.5% (detection limit 50 ppm) up to 0 to 100%, and even 99.5 to 100% in a single device. High-quality materials in the gas channel even enable the analytical devices to be used for measurements of corrosive gas mixtures. The detector unit has no contact with the measuring gas, thus allowing it to be used under harsh conditions while guaranteeing a long service life at the same time.

     

    Long lifetime

    Gas-tight separation of the electronic and physical components when using the field analyzer design with class IP65 protection and a purging gas connection guarantees a long lifetime even under harsh ambient conditions. Moreover, the field device design is also the basic variant for the optional heating feature of the analysis unit for measuring any low-boiling components and for use in hazardous zones

    Technical specifications

    Max. number of components
    1
    Components
    Oxygen
    Minimal measuring range
    0-0,5 %
    Housing
    19″ rack / field housing








    Technical advice: 03303 / 50 40 66

    Technical advice for this product